Compact, reliable and cost-effective
Put the power of versatility to work for your lab with a smart and efficient clinical chemistry
analyzer solution.
The AU480 chemistry analyzer is the ideal primary clinical chemistry analyzer for low- to
mid-volume hospitals and laboratories, or dedicated specialty chemistry or STAT analyzer
for larger laboratories. With throughput of up to 400 photometric tests per hour
(up to 800 with ISEs), increased onboard testing, reduced sample volume
and easy operation, the AU480 delivers efficiency for laboratories around the world.
Robust testing capabilities
Efficient reagent and consumable use
Convenient operational features
No Feature available.
|
Analytical System |
Fully automated, random access chemistry system with STAT capability |
|
Analytical Principle |
Spectrophotometry and potentiometry |
|
Analytical Types |
Endpoint, rate, fixed point and indirect ISE |
|
Analytical Methods |
Colorimetry, turbidimetry, latex agglutination, homogeneous EIA, indirect ISE |
|
Simultaneously Processed Analytes |
60 photometric tests + 3 ISE, 120 pre-programmed onboard tests |
|
Throughput |
400 photometric tests/hour; maximum of 800 with ISE |
|
ISE Throughput |
200 samples/hr. Maximum tests/hr: 600 if ISE only |
|
Sample Types |
Serum, plasma, urine, other |
|
Sample Feeder |
Racks with 10 samples each (barcodes on primary tubes and on racks); capacity of 80 samples; continuous loading |
|
Sample Tubes |
Primary and secondary tubes; diameter from 11.5 to 16 mm; height from 55 to 102 mm, nestled micro sample cups |
|
STAT samples |
Up to 22 positions for STAT samples via cooled turntable. Easy-to-operate interrupts between tests. Auto-repeat run capability. |
|
Sample Volume |
1–25 μl in 0.1 μl steps (1-25 μl for repeats) |
|
Reagent Capacity |
76 positions for (R1+R2, detergent position). Holds 15, 30 and 60ml bottles. |
|
Reagent Storage |
Refrigerated (4–12°C) |
|
Reagent Volume |
R1: 10–250 μl; R2: 10–250 μl (in 1 μl increment) |
|
Total Reaction Volume |
90–350 μl |
|
Reaction Cuvette |
Permanent glass cuvettes |
|
Reaction Time |
Up to 8 minutes, 37.5 seconds |
|
Reaction Incubation |
37°C, dry bath |
|
Photometry System |
Direct assay through the reaction cuvette (0–3.0 OD) mono and bichromatic measurements possible |
|
Wavelength |
13 different wavelengths between 340–800 nm |
|
Calibration |
Auto calibration, cooled calibrator positions; master calibration established by two-dimensional barcode |
|
Quality Control |
Auto QC, cooled QC positions |
|
Rerun |
1. Auto rerun and manual rerun available |
|
Sample Pre-dilution |
Dilution ratio: Fixed to 3, 5, 10, 15, 20, 25, 50, 76 or 100 |
|
Test Requisition |
Individual and profile test requisition via online, mouse, function keys and touch-screen |
|
Sample Integrity |
Lipaemia, Haemolysis and Icterus analysis. Sample clot detection and probe crash protection. |
|
Online |
RS232C. Full uni- and bidirectional communication possible. |
|
Software |
Windows 7® |
|
Data Storage |
Sample number 100,000 samples. Reaction Data 200,000 tests. |
|
Dimensions (W x H x D) |
ANL: 1450 mm x 1205 mm x 770 mm |
|
Power Supply |
100–240v; 60 Hz/ < 3.5 kVA |
Please fill out the form below to receive product details.
Simple returns policy