OPTI Medical

 

Cassettes E-Plus E - Ca E- Lytes
Tests pH, Na+, K+, Cl-, Ca++ pH, Ca++ Na+, K+, Cl-

Specifications OPTILION
Throughput Low / Medium
Sample Volume 125 µ L
Calibration Time <90 seconds
Measurement Time <120 seconds
Typical Cycle Time <210 seconds
Operating Temperature 10-32°C (50-90 °F)
Sample Type Whole blood, serum or plasma
Measurement Principle Fluorescence
Sample Application Syringe or Sample Collection
Data Storage 200 patient records
30 QC/SRC samples per level
300 user IDs and passwords

  Sample Type Available Units Measurement Range Display Resolution
Parameter Whole Blood Plasma Serum Default Other (Default Units) (Low/High)
pH X X X pH units   6.8 - 8.0 0.01 / 0.001
Na+ X X X mmol/L mg/dL 100-190 1/0.1
k+ X X X mmol/L mg/dL 1.0-9.5 0.1/0.01
Ca++ X X X mmol/L mg/dL 0.3-2.0 0.01
CI- X X X mmol/L mg/dL 65-145 1/0.1